Syringe Pump Model: KL602

Syringe Pump Model: KL602

Product Features:

1. Applicable syringe size: 10, 20, 30, 50/60 ml.

2. Automatic syringe size detection.

3. Automatic anti-bolus.

4. Drug library.

5. Wireless management: central monitoring by Infusion Management System with maximum 255 Pumps.

6. Stackable up to 4 Syringe Pumps (4-in-1 Docking Station) or 6 Syringe Pumps (6-in-1 Docking Station) with single power cord.

Product Description

Product Specifications:

Model RC-KL602
Syringe Size 10, 20, 30, 50/60 ml
Applicable Syringe Compatible with syringe of any standard
VTBI 0.1-9999 ml

<1000 ml in 0.1 ml increments

≥1000 ml in 1 ml increments

Flow Rate Syringe 10 ml: 0.1-400 ml/h

Syringe 20 ml: 0.1-600 ml/h

Syringe 30 ml: 0.1-900 ml/h

Syringe 50/60 ml: 0.1-1300 ml/h

<100 ml/h in 0.1 ml/h increments

≥100 ml/h in 1 ml/h increments

Bolus Rate Syringe 10 ml: 400 ml/h

Syringe 20 ml: 600 ml/h

Syringe 30 ml: 900 ml/h

Syringe 50/60 ml: 1300 ml/h

Anti-Bolus Automatic
Accuracy ±2% (mechanical accuracy ≤1%)
Infusion Mode Memory mode

Flow rate: ml/min, ml/h

Time-based

Body weight: mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h etc.

KVO Rate 0.1-1 ml/h (in 0.1 ml/h increments)
Alarms Occlusion, near empty, end program, low battery, end battery,

AC power off, motor malfunction, system malfunction, reminder alarm,

pressure sensor error, syringe installation error, syringe drop off

Additional Features Real-time infused volume, automatic power switching,

automatic syringe identification, mute key, purge, bolus, anti-bolus,

system memory

Drug Library Available
Occlusion Sensitivity High, medium, low
Power Supply, AC 110/230 V (optional), 50/60 Hz, 20 VA
Battery 9.6±1.6 V, rechargeable
Battery Life 7 hours at 5 ml/h
Working Temperature 5-40℃
Relative Humidity 20-90%
Atmospheric Pressure 860-1060 hpa
Size 314*167*140 mm
Weight 2.5 kg
Safety Classification Class Ⅱ, type CF