Product Description
Product Features:
| 1. CPAP ventilation modes keeps children spontaneous breath and reduces children breath work. |
| 2. Continues to provide CPAP mode of constant flow and limited pressure for newborn baby. |
| 3. Delivery air/O2 mixture with properly adjusted O2 concentration and continuous positive pressure to premature baby, newborn baby and child which will be highly improve the safety and efficiency of oxygen therapy. |
| 4. Equipped with build-in air blender, ensuring O2 concentration adjustment to be stable. |
| 5. Reliable CPAP control valve and pressure monitoring system improve CPAP adjustment precision. |
| 6. Optional parts: humidifier with heating devices. |
| 7. High cost-effective and reliable quantity. |
| 8. Easy operation, compact appearance and design let user could operate more easily. |
Product Technical Data:
|
1)Parameter adjustment |
|
|
CPAP adjusting range |
0 cmH2O ~ 10 cmH2O |
|
Continuous flow adjusting range |
1.5 L/min ~ 15 L/min |
|
Oxygen Concentration |
21%-100% |
|
2)Parameter monitoring range |
|
|
CPAP/airway pressure |
|
|
——Monitoring range |
0 cmH2O ~ 60 cmH2O |
|
——Accuracy |
±(2 % full scale + 4 % actual value) |
|
Continuous flow |
|
|
——Monitoring range |
1.5 L/min ~ 15 L/min |
|
——Accuracy |
±4 % |
|
3) Gas supply |
|
|
Gas source |
O2(For medical use), Air(For medical use) |
|
Rated working pressure |
350 kPa |
|
Output pressure of regulator |
350 kPa±50 kPa (300 kPa~400 kPa) |
Product Working environment:
| Ambient temperature | 5 ℃~ 40 ℃ |
| Relative humidity | ≤80 % |
| Atmospheric pressure | 960 hPa~1040 hPa |
| Storage and transportation environment | |
| Ambient temperature | -10 ℃~+40 ℃ |
| Relative humidity | ≤90 % |
| Atmospheric pressure | 860 hPa~1060 hPa |