Product Technical Data:

Application: Adult/Pediatric/Neonatal/Medicine/Surgery/ Operating Room/ICU/CCU

Product Description

Product Features:

1) Light & handy

2)Measurement: ECG/ SpO2/NIBP/HR&PR/Resp

3)Network: Wire/wireless network & online update

4)Triplex alarms: Light/Sound (High/medium/low levels adjustable) Electrode Disconnection Alarm (Diagram)

5)Windows XP Operating System

6)Easy operation/stable performance

7) Electrotome/defibrillation tolerance

8) Restore within 5 seconds after defibrillation

9) Power Supply: AC & DC Data Storage

10)Trend diagram/table: 720h

11)NIBP review: 10000 events

12)Wave review: 12h

13)Alarm review: 200 alarm events

14) Support drug concentration titration analysis

Physical Specification

Product dimension & weight Equipment: 370mm×160mm×350mm&4.5kg

Package: 420mm×380 mm×285mm&6.6kg

Monitor

15” real color TFT screen,   Resolution:1024*768

Battery

Electrical Power

11.1V 24wh Li-ion battery

AC: 100 ~ 240V, 50Hz/60Hz

2h operating after full charge(one piece)

DC: Built-in rechargeable battery

5min operating after low battery alarm

Product Technique Specification:

ECG

3Leads:     RA, LA, LL

Lead mode:  I, II, III

5 Leads: RA, LA, LL, RL, V

Lead mode: Ⅰ,Ⅱ, Ⅲ, AVR, AVL, AVF, V

Resp

Method: impedance between RA-LL Resp Impedance range: 0.3 ~ 3Ω

Base Impedance range: 200Ω ~ 4000Ω

Bandwidth: 0.3 ~ 2.5Hz

Battery

11.1V 24wh Li-ion battery

2h operating after full charge(one piece)

5min operating after low battery alarm

AC and Drift tolerance

Segment Range:

-2.0mV ~ +2.0mV (accuracy 0.02mV)

Heart/Pulse Rate

Measure range:

Adult 15 ~ 300 bpm Neonatal/Pediatric  15 ~ 350 bpm Accuracy: ±1%

Resolution:        1 bpm

NIBP

Method: Oscillometry

Measure mode: Manual, Auto, STAT Measure

Heart/Pulse Rate

Measure range:Adult 15 ~ 300 bpm Neonatal/Pediatric 15 ~ 350 bpm   Accuracy: ±1%

Resolution: 1 bpm

Optional: Temp

Channel 2

Measure & alarm range 0 ~ 50°C Resolution 0.1°C

Accuracy (no sensor) ± 0.1°C (0~50°C

Support heart Pacemaking Analysis

13 Arrhythmia Analysis:

Asystole, V-Fibrillation/Tachycardia, Single/Paired/Multiform PVC, Bigeminy, Trigeminy, Polycardia, P-0N-T, Bradycardia, Leaky-wave, Pacemaker

not captured, Pacemaker failure

SpO2

Measurement range: 0 ~ 100%

Resolution: 1%

Accuracy: 70% ~ 100% ±2%

40% ~ 69%      ±3%

System Inputs & Outputs

NIBP/SpO2/Temp/ECG/IBP Sensor Sockets RJ45 port

Product Standard Package:

Main unit

(15”screen) NIBP

NIBP cuff & tube

ECG cable

SpO2  sensor

Temperature probe

Lithium-ion battery

Optional Accessories : Touch screen, Trolley bracket, Hanging bracket, Monitor recorder, Vehicle-mounted power supply